Emerson MicroCEM Camera Accessories User Manual


 
Instruction Manual
748467-A
January 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-7
Model MicroCEM
3-3 MicroCEM SETTINGS
The MicroCEM Settings dialog is only
available to users with MicroCEM Settings
permission. If a user is not currently logged
in, the login dialog will be displayed. If the
current user doesn’t have MicroCEM Set-
tings permission, a message will be dis-
played which reads “Permission denied”.
When the MicroCEM Settings are invoked
from the Tools menu or the MicroCEM Set-
tings button, the MicroCEM Settings tabbed
dialog is displayed. The Range page (tab)
is displayed initially.
a. Range
The Range Settings page is used to set
the range of the Emissions analog out-
puts. The outputs support dual range
mode. When the emission is below the
Range 1 value, the output switches to
Range 1 mode and the Range 1 value
becomes the full-scale value of the out-
put. The range indication digital output
will change to the Range 1 state.
When the emission is above the Range
1 value, the output switches to Range 2
mode and the Range 2 value becomes
the full-scale value of the output. The
range indication digital output will
change to the Range 2 state. The
Range 2 settings cannot be changed
and are factory-set.
NOTE
If only one range is needed, set the
Range 1 values to the range 2 value.
This will disable the dual range fea-
ture.
Figure 3-8. Range Settings
Tabs allow selection of the
MicroCEM Settings pages