Emerson MicroCEM Camera Accessories User Manual


 
Instruction Manual
748467-A
January 2002
Rosemount Analytical Inc. A Division of Emerson Process Management Operation 3-9
Model MicroCEM
d. Manual Calibration
A dry-run Calibration may be initiated
from the Manual Calibration page of the
MicroCEM Settings. The results of the
calibration will not be applied and only
provide a dry run of the calibration. If
desired a partial calibration may be in-
voked for one or more of the emission
types. While the manual calibration is
in process, a calibration progress dialog
will be displayed as shown in Figure
3-26. When the manual calibration is
completed, the results are displayed in
the Manual Calibration Results dialog
as shown in Figure 3-12. If the Local
Calibration checkbox is checked, the
Local Calibration valve will be used
during the calibration rather than the
probe Calibration valve.
NOTE
“Start Autocalibration now” will in-
voke a real calibration and will apply
new correction factor results when
done.
Figure 3-11. Manual Calibration
Figure 3-12. Manual Calibration Results