Emerson 54e pH/ORP Photo Scanner User Manual


 
MODEL 54e pH/ORP SECTION 4.0
CALIBRATION
3. A screen like the one on the left will appear after stabilization.
Use the arrow keys to select the correct buffer.
If the correct buffer does not appear, refer to Section 5.9 and select
a buffer group that includes the desired buffers. As usual, to cancel
the calibration, press Abort (F1).
When the correct buffer is shown, remove the sensor from buffer 1,
rinse and gently dry it. Press Cont (F3) to continue.
NOTE
The calibration calculation does not occur until buffer 2 is com-
pleted. Aborting the calibration at this point will not change the
reading of the controller.
4. This screen appears immediately after pressing the Cont key in
step 3.
Place the sensor into the second buffer. Press Cont (F3) to proceed.
The word "Wait" flashes (as before) until the sensor is stabilized.
Press Abort (F1) to cancel the calibration if the process is taking too
long. (see step 2).
5. A screen like the one on the left will appear after stabilization. Use the
arrow keys to select the correct buffer, and press Cont (F3). If
the correct buffer does not appear, refer to Section 5.9 and select a
buffer group that includes the desired buffers. As usual, to cancel the
calibration, press Abort (F1).
This completes the calibration. The controller has now recalculated
the slope and the zero offset of the sensor.
CALIBRATION NOTES
1. A two-point calibration should always be followed by a sin-
gle-point calibration (see Section 4.4).
2. After a buffer calibration, the electrode slope value can be
checked to see if the electrode is aging.
3. If the sensor is not at the same temperature as the buffer the cal-
ibration will be in error.
NOTE
If hold mode was turned ON, be certain to install the sensor back
in the process and change the setting to OFF to resume normal
operation before leaving the controller. The hold mode screen
will appear again before the main display is shown. Follow the
same routine as in the Note for step 1 to turn the Hold Mode Off
and then press Exit (F1).
4.02 pH
Buf1 done: 4.01 @25°C
Abort Cont
10.02 pH
Buf2 done: 10.01 @25°C
Abort Cont
10.02 pH
Auto buffer cal done
Abort Cont
10.02 pH
Autocal buffer 2
Abort Cont
17