3. Menu Item Options
Output 1 units mA/%
Output 2 units mA/%
Language English/Français/Español/Deutsch/Italiano
Press the arrow key three times to access this screen. The current out-
puts can be displayed as milliamps or as percent of full scale. The default
is mA. To change any of these items, use the arrow key to highlight the
desired item and press Edit (F4). Use the arrow keys to make the change
and press Save (F4) to enter the change into memory.
Further menu items are available by pressing the arrow key repeatedly.
When the display is highlighting the item on the third line, the end of the
menu has been reached. To back up within the menu, use the up arrow
key.
MODEL 54e pH/ORP SECTION 5.0
SOFTWARE CONFIGURATION
5.5 CHOOSING DISPLAY OPTIONS
This section describes the options available for the changing of engineering
units and variables on the main display.
1. From the main menu, move the cursor down to "Program" and press
Enter 4. From the program menu, move the cursor down using the
arrow key to highlight "Configure" and press Enter 4.
The first configuration menu is displayed. With the cursor on "Display"
press Enter 4.
2. Menu Item Options
Measure pH/ORP/Redox
Resolution (pH only) 0.01pH/0.1pH
Temperature units °C/°F
The values now being used by the controller are displayed. To change
any of these items, use the arrow key to highlight the desired item and
press Edit(F4). Use the arrow keys to make the change and press Save
(F4) to enter the change into memory.
WARNING
Changing the measurement selection from pH to ORP or
ORP to pH will reset all controller settings to factory values.
Press Abort (F1) to cancel the change. Press Cont (F3) to
change the measurement. Refer to Appendix A for more
about ORP measurements.
Display
Outputs
Alarms
Exit Enter
Measure: pH
Resolution: 0.01
Temp Units: °C
Exit Enter
Measure: pH
Resolution: 0.01
Temp Units: °C
Exit Enter
Output 1: mA
Output 2: mA
Language: English
Exit Edit
Display left: AL1
Display right: AL2
Display contrast: 5
Exit Edit
WARNING: This
change resets analyzer
to factory settings
Abort Cont
31