28 PIXIS System Manual Version 2.C
Notes:
1. Make sure that there are no kinks in the hoses that impede the coolant flow. Lack
of sufficient flow can seriously harm the detector and any resulting damage is not
covered under warranty.
2. Damage caused by water leaking into the PIXIS voids the warranty.
5. Unscrew the reservoir cap (on top of the CoolCUBE
II
) and make
sure that the coolant reservoir contains coolant. If additional
coolant is required, fill with a 50:50 mixture of water and ethylene
glycol.
6. Screw the reservoir cap back in.
7. Plug the circulator into a 100-240 VAC, 47-63 Hz power source.
8. Turn the circulator on. Make sure there are no leaks or air bubbles in the hoses.
Note: Small air bubbles (about the size of bubbles in soda) are common in the
CoolCUBE
II
especially at start up and do not prevent proper operation.
If there are no problems, continue to Step 9.
If there are leaks or air bubbles, turn the circulator off and correct the problem(s)
by securing the hoses or adding more coolant to the reservoir. Turn the circulator
back on. Recheck and if there are no problems, continue to Step 9.
9. Turn the camera on.
10. Start the application software.