86 PIXIS System Manual Version 2.C
Controller Is Not Responding
If this message pops up when you click on OK after selecting the "Interface Type" during
Hardware Setup (under the WinX Setup menu), the system has not been able to
communicate with the camera. Check to see if camera has been turned ON and if the
USB 2.0 interface card, its driver, and the USB cable have been installed.
If the camera is ON, the problem may be with the USB 2 card, its driver, or the
cable connections.
If the interface card is not installed, close the WinX application and turn the
camera OFF. Follow the interface card installation instructions in provided with
your interface card, and cable the card to the USB 2 port on the rear of the
camera. Then do a "Custom" installation of the WinX application with the
appropriate interface component selected. Be sure to deselect the interface
component that does not apply to your system.
If the interface card is installed in the computer and is cabled to the USB 2 port
on the rear of the camera, close the application and turn the camera OFF. Check
the cable connections.
If the interface card was installed after the application program was installed,
close the application program and do a "Custom" installation of the WinX
application with the appropriate interface component selected. Be sure to deselect
the interface component that does not apply to your system.
CoolCUBE
II
: Low Coolant (Air in the Hoses)
If more than two inches (50.8 mm) of the coolant line is filled with air, the pump will
stop working and may be damaged. If flow stops while the pump is on, turn off the
CoolCUBE
II
and add coolant.
1. Unscrew the reservoir cap (on top of the CoolCUBE
II
) and make sure that the coolant
reservoir contains coolant. If additional coolant is required, fill with a 50:50 mixture
of water and ethylene glycol.
2. Screw the reservoir cap back in.
3. Make sure the power switch is turned off before plugging the circulator in.
4. Plug the circulator into a 100-240 VAC, 47-63 Hz power source.
5. Turn the circulator on. Make sure there are no leaks or air bubbles in the hoses.
Note: Small air bubbles (about the size of bubbles in soda) are common in the
CoolCUBEII especially at start up and do not prevent proper operation.
If there are no problems, continue to Step 6.
If there are leaks or air bubbles, turn the circulator off and correct the problem(s)
by securing the hoses or adding more coolant to the reservoir. Turn the circulator
back on. Recheck and if there are no problems, continue to Step 6.
6. Turn the camera on.
7. Start the application software.
WARNING!