Nikon DS-5MC Digital Camera User Manual


 
VIII Troubleshooting
2 Image Output
- 159 -
Symptom Corrective action
The color is poor.
Set the white balance.
=> See “2 Photographing Images” in Chapter VI.
Apply a moderate amount of color compensation.
=> See “2 Photographing Images” in Chapter VI.
Turn off hue adjustment.
=> See “2 Photographing Images” in Chapter VI.
Set the color temperature of the internal monitor and external monitor
to an appropriate level.
=> See “1.4.4 Adjusting Monitor Output for an External Monitor” in
Chapter VI.
Set the color temperature of the illumination to an appropriate level.
The contrast is low.
Set the tone appropriately.
=> See “2 Photographing Images” in Chapter VI.
Avoid adjusting the black level to the positive side of SFT.
=> See “2 Photographing Images” in Chapter VI.
Adjust the size of the field diaphragm so that it circumscribes the
viewfield.
Adjust the size of the aperture diaphragm to approximately 70% to
80% of the numerical aperture of the objective.
=> See “Chapter V Microscope Adjustment.”
Use an appropriate filter.
When using a phase contrast microscope or an interference
microscope (two luminous fluxes or multiple luminous fluxes), insert
the specified filter. With a black and white image, contrast is improved
by inserting a green interference filter (GIF). Contrast in color images
is enhanced by inserting a filter having a complementary color to that
of the specimen.
=> See “Chapter V Microscope Adjustment.”
Adjust the ambient light around the microscope.
Make the room darker, or place caps on the binocular eyepiece, for
example.
Check the contrast of the specimen itself.
Noise appears at the
edges of the menu
when shown on an
LCD monitor or
projector.
Adjust the clock phase of the LCD monitor or projector.
=> See “1.4.4 Adjusting Monitor Output for an External Monitor” in
Chapter VI.