Teledyne MODEL T703 Photo Scanner User Manual


 
Teledyne API T703 Calibrator Operation Manual Operating the T703 Calibrator
99
6.12.3. LEVELS
A LEVEL is a combination of several parameters:
An ID number for the LEVEL
An action, (e.g. GENERATE, STANDBY)
A target concentration value
An output flow rate (if applicable)
Configuration for one or both of two status output blocks.
Up to twenty levels can be defined and used with the T703 using a range of ID numbers from 0-98. Level 99 is
reserved for standby. Are not time based and do not include characteristics such as start time or duration,
therefore a single LEVEL can not switch between different concentration levels and flow rates. Separate flow
and concentration outputs must be programmed into separate LEVELs which are then individually started and
stopped either by an operator at the calibrator’s front panel or through a serial data operation over the RS-232 or
Ethernet ports.
6.12.4. ACTIVATING AN EXISTING LEVEL
To activate an existing defined LEVEL, press:
07223B DCN6378