6.1. TEST FUNCTIONS
A variety of TEST FUNCTIONS are available for viewing at the front panel whenever the calibrator is at the
MAIN MENU. These measurements provide information about the present operating status of the calibrator and
are useful during troubleshooting (see Section 11). Table 6-1 lists the available TEST functions.
To view these TEST functions, press <TST or TST> from the main menu and the Test Functions will scroll at
the top center of the display.
Table 6-1: Test Functions Defined
TEST MEASUREMENT DESCRIPTION
ACT=GENERATE XX PPB O3 Actual concentration being generated, computed from
real-time inputs.
TARG=GENERATE XX PPB O3
Target concentration to generate.
OUT FLW=X.XXX LPM
Output flow rate (computed from regulator pressure).
REG PRESS=XX.X PSIG
Regulator pressure.
BOX TMP=XX.X ºC
Internal chassis temperature.
OGEN REF=XXXX.X MV
O
3
generator reference detector reading.
O3GENDRV=XXXX.X MV
O
3
generator lamp drive output.
O3LAMPTMP=XX.X ºC
O
3
generator lamp temperature.
PH MEAS=XXXX.X MV
Photometer detector measure reading.
PH REF=XXXX.X MV
Photometer detector reference reading.
PH FLW=X.XXX LPM
Photometer sample flow rate.
PH LTEMP=XX.X ºC
Photometer lamp temperature.
PH SPRES=XX.X IN-HG-A
Photometer sample pressure.
PH STEMP=XX.X ºC
Photometer sample temperature.
PH SLOPE=X.XXX Photometer slope computed during zero/span bench
calibration.
PH OFFST=X.X PPB Photometer offset computed during zero/span bench
calibration.
PHOTO STABIL=X.X PPB Photometer concentration stability (standard deviation of
25 bench concentration samples taken 10 seconds apart).
TEST=XXXX.X MV Value output to TEST_OUTPUT analog output, selected
with TEST_CHAN_ID variable.
TIME=HH:MM:SS
Current instrument time of day clock. (24 hour format)