3.4.4. FUNCTIONAL CHECK
1. After the calibrator’s components have warmed up for at least 30 minutes, verify that the software
properly supports any hardware options that are installed.
2. Check to make sure that the calibrator is functioning within allowable operating parameters. Appendix C
includes a list of test functions viewable from the calibrator’s front panel as well as their expected values.
These functions are also useful tools for diagnosing problems with your calibrator (see Section7.5.2).
The enclosed Final Test and Validation Data sheet (part number 05760) lists these values before the
instrument left the factory.
To view the current values of these parameters press the following touch screen button sequence on the
calibrator’s front panel. Remember until the unit has completed its warm up these parameters may not
have stabilized.
3. If your calibrator is operating via Ethernet and your network is running a dynamic host configuration
protocol (DHCP) software package, the Ethernet option will automatically configure its interface with
your LAN. However, it is a good idea to check these settings to make sure that the DHCP has
successfully downloaded the appropriate network settings from your network server (See Section 7.5.2).
If your network is not running DHCP, you will have to configure the calibrator’s interface manually (See
Section 7.5.1.1.