Teledyne MODEL T703 Photo Scanner User Manual


 
Teledyne API T703 Calibrator Operation Manual Getting Started
35
3.2.5. CONNECTING THE CONTROL OUTPUTS
The calibrator is equipped with 12 opto-isolated, digital control outputs. These outputs are activated by the
T703’s user-programmable, calibration sequences (see Section 6.5.1.6 for instructions on assigning the control
outputs to specific
calibration sequences)
These outputs may be used to interface with devices that accept logic-level digital inputs, such as programmable
logic controllers (PLCs), dataloggers, or digital relays/valve drivers.
They are accessed via a 14-pin connector on the calibrator’s rear panel (see Figure 3-4).
CONTROL OUTPUTS
1 2 3 4 5 6 7 8 9 10 11 12 E
Figure 3-10: T703 Digital Control Output Connector
NOTE
Most PLCs have internal provisions for limiting the current the input will draw. When connecting to a
unit that does not have this feature, external resistors must be used to limit the current through the
individual transistor outputs to 50mA (120 for 5V supply).
The pin assignments for the control outputs are:
Table 3-5: T703 Control Input Pin Assignments
PIN # STATUS DEFINITION CONDITION
1 - 12 Outputs 1 through 12 respectively Closed if the sequence or sequence step activating output is operating
E Emitter BUSS The emitters of the transistors on pins 1 to 8 are bussed together.
Digital Ground The ground level from the calibrator’s internal DC power supplies.
07223B DCN6378