Teledyne MODEL T703 Photo Scanner User Manual


 
Teledyne API T703 Calibrator Operation Manual Specifications and Approvals
21
2. SPECIFICATIONS AND APPROVALS
2.1. SPECIFICATIONS
Table 2-1: T703 System Specifications
Linearity
+/- 1.0% of full scale
Precision
1.0 ppb
Response Time 180 seconds to 95%
Stability (7-days) 1% photometer feedback; 3% without photometer feedback (CNST or REF)
Temperature Range 5-40ºC
Humidity Range 0 - 95% RH, non-condensing
Operating Altitude 10,000 ft Maximum
Dimensions (H x W x D) 7” (178 mm) x 17” (432 mm) x 24” (609 mm)
Weight 35.5 lbs (16.1 kg) including internal zero air pump
AC Power 100V – 120V, 220V – 240V, 50/60Hz
Analog Output Ranges
0.1 V, 1 V, 5 V or 10 V
Range with 5% under/over-range
Analog Output Resolution 1 part in 4096 of selected full-scale voltage (12 bit)
Standard I/O
1 Ethernet: 10/100Base-T
2 RS-232 (300 – 115,200 baud)
2 USB device ports
8 opto-isolated digital control outputs
12 opto-isolated digital control inputs
8 opto-isolated digital status outputs
1 analog output, user configurable
Optional I/O
1 USB com port
1 RS485
8 analog inputs (0-10V, 12-bit)
4 digital alarm outputs
Multidrop RS232
3 4-20mA current outputs
Certifications
EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section 15.107 Class
A, ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548 (w/A1 & A2; 97)
Class A.
IEC 61010-1:90 + A1:92 + A2:95,
07223B DCN6378