Teledyne MODEL T703 Photo Scanner User Manual


 
Appendix C
Warranty/Repair Questionnaire
T703, M703E
(05747C DCN5798)
TELEDYNE API CUSTOMER SERVICE
Email: api-customerservice@teledyne.com
PHONE: (858) 657-9800 TOLL FREE: (800) 324-5190 FAX: (858) 657-9816
C-1
CUSTOMER:_______________________________ PHONE: _____________________________________
CONTACT NAME: __________________________ FAX NO. _____________________________________
SITE ADDRESS:____________________________________________________________________________
MODEL TYPE: ______________ SERIAL NO.:_________________ FIRMWARE REVISION:_____________
Are there any failure messages? _______________________________________________________________
_________________________________________________________________________________________
__________________________________________________________________________________________
__________________________________________________________________________________________
________________________________________________________________________
(Continue on back if necessary)
PLEASE COMPLETE THE FOLLOWING TABLE:
PARAMETER
Name in E-Series software
versions prior to v.C.1
Name in T-Series and
in E-Series w/software
v. C.1 and higher.
RECORDED VALUE
ACCEPTABLE
VALUE
ACT ACT
PPB
1% OF TARG
TARG TAR
PPB
50 – 1000 PPB
OUTPUT FLOW OUT FLW
LPM
2 – 5 LPM
REG PRESSURE REG PRES
PSIG
15 ± 2 PSIG @ 5 LPM
BOX TEMP BOX TMP
ºC
20 – 35 ºC
O3 GEN REF
1
O3GEN REF
1
mV
0 – 5000 mV
O3 GEN DRIVE
1
O3GENDRV
1
mV
0 – 5000mV
O3 LAMP TEMP
1
O3LAMPTMP
1
ºC
48 ± 1 ºC
PHOTO MEASURE
2
PH MEAS
2
mV
2500 – 4700 mV
PHOTO REFERENCE
2
PH REF
2
mV
2500 – 4700 mV
PHOTO FLOW
2
PH FLW
2
LPM
0.720 – 0.880 LPM
PHOTO LAMP TEMP
2
PH LTEMP
2
ºC
58 ± 1 ºC
PHOTO SPRESS
2
PH SPRES
2
IN-HG-A
-1” AMBIENT IN-HG-A
PHOTO STEMP
2
PH STEMP
2
ºC
25 – 48ºC
PHOTO SLOPE
2
PH SLOPE
2
1 ± 0.15
PHOTO OFFSET
2
PH OFFST
2
PPB
0 ± 10 PPB
Depending on options installed, not all test parameters shown will be available in your calibrator.
1
If ozone generator option installed.
2
If photometer option installed.
What is measured photometer flow rate ____________________________________________________ cc
3
/min
What is measured O
3
generator flow rate? _________________________________________________ cc
3
/min