Teledyne MODEL T703 Photo Scanner User Manual


 
Operating The T703 over the Serial I/O Ports Teledyne API T703 Calibrator Operation Manual
112
7.1.7. TERMINAL OPERATING MODES
The T703 can be remotely configured, calibrated or queried for stored data through the serial ports. As
terminals and computers use different communication schemes, the calibrator supports two communicate modes
specifically designed to interface with these two types of devices.
Computer mode is used when the calibrator is connected to a computer with a dedicated interface
program.
Interactive mode is used with a terminal emulation programs such as HyperTerminal or a “dumb”
computer terminal. The commands that are used to operate the calibrator in this mode are listed in Table
7-2.
7.1.7.1. Help Commands in Terminal Mode
Table 7-2: Terminal Mode Software Commands
COMMAND Function
Control-T Switches the calibrator to terminal mode
(echo, edit). If mode flags 1 & 2 are OFF,
the interface can be used in interactive
mode with a terminal emulation program.
Control-C Switches the calibrator to computer mode
(no echo, no edit).
CR
(carriage return)
A carriage return is required after each
command line is typed into the
terminal/computer. The command will not
be sent to the calibrator to be executed until
this is done. On personal computers, this is
achieved by pressing the ENTER key.
BS
(backspace)
Erases one character to the left of the
cursor location.
ESC
(escape)
Erases the entire command line.
? [ID] CR This command prints a complete list of
available commands along with the
definitions of their functionality to the
display device of the terminal or computer
being used. The ID number of the
calibrator is only necessary if multiple
calibrators are on the same
communications line, such as the multi-
drop setup.
Control-C
Pauses the listing of commands.
Control-P
Restarts the listing of commands.
07223B DCN6378