Teledyne MODEL T703 Photo Scanner User Manual


 
Teledyne API T703 Calibrator Operation Manual T703 Calibration and Verification
141
8.2.6. PERFORMING AN AUTOMATIC CALIBRATION OF THE O
3
GENERATOR
To run the automatic O
3
generator calibration program, press:
Test runs automatically
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ
Make sure that the
calibrator is in standby
mode.
SETUP X.X
O3 SEQ CFG CLK PASS EXIT
SETUP X.X
COMM VARS EXIT
SETUP X.X
EXIT
Toggle to enter the correct
DIAG
PREV ENTR EXIT
DIAG
PREV NEXT EXIT
Continue pressing until ...
DIAG
PNTS EXIT
DIAG
DIAG
aborts
the calibration
07223B DCN6378