Teledyne API T703 Calibrator Operation Manual Operating The T703 over the Serial I/O Ports
111
7.1.6. MACHINE ID
Each type of Teledyne API calibrator is configured with a default ID code. The default ID code for all T703
calibrators is 703. The ID number is only important if more than one calibrator is connected to the same
communications channel such as when several calibrators are on the same Ethernet LAN, in a RS-232 multidrop
chain (See Section 7.3) or operating over a RS-485 n
e
twork (See Section 7.4). If two calibrators of the same
model type are used on one channel, their ID codes must differ; opt to change one or both.
To edit the instrument’s ID code, press:
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
SETUP X.X COMMUNICATIONS MENU
ID INET COM1 COM2 EXIT
SETUP X.X MACHINE ID:703 ID
0 7 0 0 ENTR EXIT
Toggle to cycle
through the available
character set: 0-9
EXIT discards the new
setting
ENTR accepts the
new setting
The ID number is only important if more than one calibrator is connected to the same communications channel
(e.g., a multi-drop setup). Different models of Teledyne API’s calibrators have different default ID numbers, but if
two calibrators of the same model type are used on one channel (for example, two T703’s), the ID of one
instrument needs to be changed.
The ID can also be used for to identify any one of several calibrators attached to the same network but situated
in different physical locations.