Teledyne 703E Photo Scanner User Manual


 
TELEDYNE API
Operating The M703E over the Serial I/O Ports M703E Calibrator Operator’s Manual
7.1.7.5. COM Port Password Security
In order to provide security for remote access of the M703E, a LOGON feature can be enabled to require a
password before the instrument will accept commands. This is done by turning on the
SECURITY MODE (Mode
4, Section 7.1.4). Once the
SECURITY MODE is enabled, the following items apply.
A password is required before the port will respond or pass on commands.
If the port is inactive for one hour, it will automatically logoff, which can also be achieved with the
LOGOFF command.
Three unsuccessful attempts to log on with an incorrect password will cause subsequent logins to be
disabled for 1 hour, even if the correct password is used.
If not logged on, the only active command is the '?' request for the help screen.
The following messages will be returned at logon:
LOGON SUCCESSFUL - Correct password given
LOGON FAILED - Password not given or incorrect
LOGOFF SUCCESSFUL - Connection terminated successfully
To log on to the M703E calibrator with
SECURITY MODE feature enabled, type:
LOGON 940331
940331 is the default password. To change the default password, use the variable RS232_PASS issued as
follows:
V RS232_PASS=NNNNNN
Where N is any numeral between 0 and 9.
98 05744 Rev B