M703E Calibration and Verification M703E Calibrator Operator’s Manual
8.4.2. CALIBRATING THE OUTPUT GAS FLOW
8.4.2.1. Output Gas Flow Set Up
The procedure described in this section requires an independent, calibrated flow meter/monitor and the following
set up:
O
3
FLOW
SENSOR
O
3
Generator Assembly
O
3
GENERATOR
Flow Control
(5.0 lpm)
CHARCOAL
SCRUBBER
Flow Control
(1.0 LPM)
CAP
CAP
CAP
Flow Control
(100 cm
3
/min)
Figure 8-5: Output Flow Calibration Monitor Point
130 05744 Rev B