M703E Calibrator Operator’s Manual M703E Calibration and Verification
8.2.5. TURNING O
3
GENERATOR CALIBRATION POINTS ON / OFF
To enable or disable an existing calibration point, press:
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
Make sure that the M703E
is in standby mode.
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
SETUP X.X ENTER PASSWORD
818 ENTREXIT
Toggle these keys to enter
the correct PASSWORD
DIAG SIGNAL I/O
PREV NEXT ENTR EXIT
DIAG O3 GEN CALIBRATION
PREV NEXT ENTR EXIT
Continue pressing NEXT until ...
DIAG O3 GEN CALIBRATION
CAL PNTS EXIT
DIAG O3GEN 1) 500 MV, 5.0 MIN, ON
PREV NEXT INS DEL EDIT PRNT EXIT
Continue pressing PREV & NEXT until your
reach the point to be turned ON/OFF
DIAG O3GEN 8) 1500 MV, 5.0 MIN, ON
PREV NEXT INS DEL EDIT PRNT EXIT
DIAG O3GEN CAL. POINT DRIVE:0 MV
<SET SET> EDIT EXIT
DIAG O3GEN CAL. POINT ENABLELD:ON
<SET SET> EDIT EXIT
Continue pressing SET> until ...
DIAG O3GEN CAL. POINT ENABLELD:ON
ON ENTR EXIT
Toggle this key to turn the
point ON / OFF
EXIT discards
the new setting
ENTR accepts
the new setting
05744 Rev B 123