APPENDIX A-2: Setup Variables For Serial I/O, Software Version C.0 M703E Calibrator Operator’s Manual
A-12 05745 Rev C
M703E SETUP VARIABLES FOR LATEST REVISION
SETUP VARIABLE
NUMERI
C UNITS
DEFAULT
VALUE
VALUE RANGE DESCRIPTION
O3_PUMP_STARTUP
— ON OFF, ON
O
3
pump startup enable. ON enables startup
procedure.
O3_PUMP_MIN_FLOW LPM 0.2 0–1
Minimum flow rate that indicates O
3
pump is
on.
O3_PUMP_TIMEOUT
Seconds 30 1–180 O
3
pump startup timeout.
O3_PUMP_PULSE
Seconds 0.5 0.1–10 O
3
pump power off pulse duration.
PHOTO_CYCLE
Seconds 10 0.5–30 Photometer lamp temperature control cycle period.
PHOTO_PROP
— 0.5 0–10
Photometer lamp temperature PID proportional
coefficient.
PHOTO_INTEG
— 0.05 0–10
Photometer lamp temperature PID integral
coefficient.
PHOTO_DERIV
— 0.2 0–10
Photometer lamp temperature PID derivative
coefficient.
PHOTO_FLOW_SLOPE
1 0.001–100 Slope term to correct photometer sample flow rate.
O3_DEF_DRIVE
mV 0 0–5000 O
3
generator default drive setting.
O3_GEN_MODE
— BENCH
CNST,
REF,
BNCH
O
3
generator control mode. Enclose value in double
quotes (") when setting from the RS-232 interface.
O3_MIN_CONC
PPB 25 0–100
O
3
generator minimum reliable concentration. Less
than this is treated as zero.
REF_DELAY
Seconds 60 1–300 O
3
generator reference feedback control delay.
REF_FREQ
Seconds 1 1–60 O
3
generator reference adjustment frequency.
REF_FSIZE
Samples 4 1–10 O
3
generator reference filter size.
REF_INTEG
— 0.1 0–10 O
3
generator reference PID integral coefficient.
REF_DERIV
— 0.2 0–10 O
3
generator reference PID derivative coefficient.
BENCH_DELAY
Seconds 120 1–300 O
3
generator bench feedback control delay.
BENCH_FREQ
Seconds 10 1–60 O
3
generator bench adjustment frequency.
BENCH_FSIZE
Samples 3 1–10 O
3
generator bench filter size.
BENCH_INTEG
— 0.2 0–10 O
3
generator bench PID integral coefficient.
BENCH_DERIV
— 0.5 0–10 O
3
generator bench PID derivative coefficient.
DRIVE_STABIL
mV 10 0.1–100
O
3
generator drive stability limit to update
concentration cache.
CACHE_RESOL
PPB 2 0.1–20
O
3
generator cache un-normalized concentration
resolution.
O3_LAMP_CYCLE
Seconds 2 0.5–30 O
3
generator lamp temperature control cycle period.
O3_LAMP_PROP
1/DegC 0.2 0–10
O
3
generator lamp temperature PID proportional
coefficient.
O3_LAMP_INTEG
Gain 0.01 0–10
O
3
generator lamp temperature PID integral
coefficient.
O3_LAMP_DERIV
Gain 0.2 0–10
O
3
generator lamp temperature PID derivative
coefficient.
OUTPUT_FLOW_SLOPE — 1 0.001–100
Slope term to correct output flow rate.