Teledyne 703E Photo Scanner User Manual


 
TELEDYNE API
M703E Calibrator Operator’s Manual Operating the M703E Calibrator
6.5.2.1. The Generate Step
This step operates and is programmed similarly to the GENERATE AUTO.
At the end of the programming sequence, the M703E firmware will automatically insert a
DURATION step that
needs to be defined.
To insert a
GENERATE step into a sequence, press:
INSERT STEP Submenu
SETUP X.X INSERT STEP: GENERATE
PREV NEXT ENTR EXIT
Starting at the INSERT STEPS
Submenu
EXIT discards the new
target concentration
ENTR accepts the new
target concentration
SETUP X.X GENERATE:0.0 PPB O3
0.000PCTO3ENTREXIT
SETUP X.X GENERATE:ZERO
ZERO ENTR EXIT
SETUP X.X 3) DURATION : 10.0 MIN
PREV NEXT INS DEL EDIT EXIT
Toggle this key to
switch between
ZERO AIR and O
3
modes.
SETUP X.X GENERATE:0.0 PPB O3
0 0 0 .0 PPB O3 ENTR EXIT
Toggle these keys
to set the target
concentration.
Toggle this key to
to scroll through the
available units of
measure
SETUP X.X DURATION: 1.0 MIN
001.0 ENTREXIT
Toggle these keys
to set DURATION
of this step
05744 Rev B 51