M703E Calibrator Operator’s Manual Operating the M703E Calibrator
05744 Rev B 39
6.4. GENERAL INFORMATION ABOUT THE GENERATE MODE
The GENERATE mode is the mode of operation where the M703E is actively producing calibration gas, either
zero or some specified concentration of ozone. In the
GENERATE mode the Zero Air Pump (if enabled) and
Photometer Pump are turned on.
6.4.1. GENERATE AUTO: Basic Generation of Calibration Gas
This is the simplest procedure for generating calibration gas mixtures. To generate calibration gas, press
The M703E will now enter
GENERATE mode.