6.9.2. TEST CHANNEL CALIBRATION
TEST CHANNEL calibration needs to be carried out on first startup of the analyzer (performed in the factory as
part of the configuration process) or whenever re-calibration is required. The analog outputs can be calibrated
automatically or adjusted manually.
During automatic calibration, the analyzer tells the output circuitry to generate a zero mV signal and high-scale
point signal (usually about 90% of chosen analog signal scale) then measures actual signal of the output. Any
error at zero or high-scale is corrected with a slope and offset.
Automatic calibration can be performed via the
AOUTS CALIBRATION command, or by using the CAL button
located inside
TEST_CHANNEL submenu. By default, the analyzer is configured so that calibration of TEST
CHANNEL
can be initiated with the AOUTS CALIBRATION command.
6.9.2.1. Enabling or disabling the TEST CHANNEL Auto-Cal Feature
To enable or disable the Auto-Cal feature for the TEST CHANNEL, press.
From the
AIO CONFIGURATION SUBMENU
(See Section 6.9.1.1.)
DIAG ANALOG I/O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG AIO TEST_OUTPUT: 5V, OVR, NOCAL
<SET SET> EDIT EXIT
DIAG AIO TEST_OUTPUT: AUTO CAL.:ON
<SET SET> EDIT EXIT
DIAG AIO TEST_OUTPUT: AUTO CAL.:ON
ON ENTR EXIT
Toggle this key to
turn AUTO CAL
ON or OFF
(OFF = manual
calibration mode).
ENTR accepts
the new setting.
EXIT ignores the
new setting
Continue pressing SET> until ...
DIAG AIO TEST_OUTPUT: RANGE: 5V
SET> EDIT EXIT
DIAG AIO TEST_OUTPUT: AUTO CAL.:OFF
OFF ENTR EXIT
Continue pressing SET> until you reach the
output to be configured
DIAG AIO AOUTS CALIBRATED: NO
SET> CAL EXIT
NOTE:
TEST CHANNELS
configured for 0.1V full
scale should always be
calibrated manually.