3.4.4. FUNCTIONAL CHECK
7. After the calibrator’s components have warmed up for at least 30 minutes, verify that the software
properly supports any hardware options that are installed.
8. Check to make sure that the calibrator is functioning within allowable operating parameters. Appendix C
includes a list of test functions viewable from the calibrator’s front panel as well as their expected values.
These functions are also useful tools for diagnosing problems with your calibrator (see Section7.5.2).
The enclosed Final Test and Validation Data sheet (part number 05760) lists these values before the
instrument left the factory.
To view the current values of these parameters press the following key sequence on the calibrator’s front
panel. Remember until the unit has completed its warm up these parameters may not have stabilized.
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ
2
MSG CLR
1
SETUP
Toggle <TST TST> keys
to scroll through list of
functions
ACT=GENERATE [Value] PPB O3
TARG=GENERATE [Value] PPB O3
OUTPUT FLOW=[Value] LPM
REG PRESSURE=[Value] PSIG
BOX TEMP=[Value]ºC
O3 GEN REF=[Value] MV
O3 GEN DRIVE=[Value] MV
O3 LAMP TEMP=[Value]ºC
PHOTO MEASURE[Value] MV
PHOTO FLOW=[Value] LPM
PHOTO LAMP TEMP=[Value]ºC
PHOTO SPRESS=[Value] IN-HG-A
PHOTO STEMP=[Value]ºC
PHOTO SLOPE=[Value]
PHOTO OFFSET=[Value] PPB
PHOTO STABIL=[Value] PPB
TEST=[Value]MV
TIME=[HH:MM:SS]
1
Only appears when warning
messages are active.
2
Only appears when an one or
more calibration sequences are
programmed into the
calibrator’s memory.
9. If your calibrator has an Ethernet card (Option 63) installed and your network is running a dynamic host
configuration protocol (DHCP) software package, the Ethernet option will automatically configure its
interface with your LAN. However, it is a good idea to check these settings to make sure that the DHCP
has successfully downloaded the appropriate network settings from your network server (See Section
7.5.2).
If your network is not running DHCP, you will have to configure the calibrator’s interface manually (See
Section 7.5.2.