Operating the M703E Calibrator M703E Calibrator Operator’s Manual
6.1. TEST FUNCTIONS
A variety of TEST FUNCTIONS are available for viewing at the front panel whenever the calibrator is at the
MAIN MENU. These measurements provide information about the present operating status of the calibrator and
are useful during troubleshooting (see Chapter 11). Table 6-1 lists the available TEST functions.
To view these TEST functions, press <TST or TST> from the main menu and the Test Functions will scroll at
the top center of the display.
Table 6-1: Test Functions Defined
TEST MEASUREMENT DESCRIPTION
ACT=GENERATE XX PPB O3 Actual concentration being generated, computed from
real-time inputs.
TARG=GENERATE XX PPB O3
Target concentration to generate.
OUTPUT FLOW=X.XXX LPM
Output flow rate (computed from regulator pressure).
REG PRESSURE=XX.X PSIG
Regulator pressure.
BOX TEMP=XX.X ºC
Internal chassis temperature.
O3 GEN REF=XXXX.X MV
O
3
generator reference detector reading.
O3 GEN DRIVE=XXXX.X MV
O
3
generator lamp drive output.
O3 LAMP TEMP=XX.X ºC
O
3
generator lamp temperature.
PHOTO MEASURE=XXXX.X MV
Photometer detector measure reading.
PHOTO REFERENCE=XXXX.X MV
Photometer detector reference reading.
PHOTO FLOW=X.XXX LPM
Photometer sample flow rate.
PHOTO LAMP TEMP=XX.X ºC
Photometer lamp temperature.
PHOTO SPRESS=XX.X IN-HG-A
Photometer sample pressure.
PHOTO STEMP=XX.X ºC
Photometer sample temperature.
PHOTO SLOPE=X.XXX Photometer slope computed during zero/span bench
calibration.
PHOTO OFFSET=X.X PPB Photometer offset computed during zero/span bench
calibration.
PHOTO STABIL=X.X PPB Photometer concentration stability (standard deviation of
25 bench concentration samples taken 10 seconds apart).
TEST=XXXX.X MV Value output to TEST_OUTPUT analog output, selected
with TEST_CHAN_ID variable.
TIME=HH:MM:SS
Current instrument time of day clock. (24 hour format)