TELEDYNE API
Operating the M703E Calibrator M703E Calibrator Operator’s Manual
6.12.5.1. Creating a GENERATE LEVEL
To create a LEVEL using the M703E’s AUTO generation function, press:
SETUP X.X GENERATE:0.0 PPB O3
0.000PCTO3ENTR EXIT
SETUP X.X GENERATE:ZERO
ZERO ENTR EXIT
Toggle this key to scroll
through the available
gas types (as
programmed during
initial setup.
SETUP X.X GENERATE:0.0 PPB O3
0 0 0 .0 PPB O3 ENTR EXIT
Toggle this key to
to scroll through the
available units of
measure
EXIT discards the
new setting
Toggle these keys
to set the target
concentration.
STANDBY LEVEL:0
0 0 ENTR EXIT
Toggle these keys until
the designation of the
existing defined level
program is reached.
EXIT discards the new
LEVEL number
ENTR accepts the new
LEVEL number
CHOOSE ACTION Submenu
SETUP X.X ACTION TO PERFORM:GENERATE
PREV NEXT ENTR EXIT
Starting at the CHOOSE ACTION Submenu
82 05744 Rev B