Teledyne 703E Photo Scanner User Manual


 
TELEDYNE API
M703E Calibrator Operator’s Manual M703E Calibration and Verification
8.4.1. CALIBRATING THE PHOTOMETER’S SAMPLE GAS FLOW
NOTE
The procedure described in this section requires an independent, calibrated gas flow meter/monitor be
connected to the EXHAUST fitting on the back of the M703E.
During the PHOTO FLOW calibration, the M703E software automatically turns the DC pump downstream from
the photometer
ON.
To perform a
PHOTO FLOW calibration, press:
05744 Rev B 129