Teledyne 703E Photo Scanner User Manual


 
TELEDYNE API
M703E Calibrator Operator’s Manual Operating the M703E Calibrator
6.5.1.7. Setting the PROGRESS Reporting Mode for the Sequences
As sequences run, the M703E calibrator reports progress by displaying a message in the MODE field of the front
panel display (See Figure 3-1). There are several types of report modes available
Table 6-5: Sequence Progress Reporting Mode
MODE
DESCRIPTION
STEP
Shows the progress as the sequence name and step number. This is the traditional display.
Example: “SO2_Test-2”.
PCT
Shows the progress as a percent (0–100%) of the total sequence duration.
Example: “SEQ 48%”
ELAP
Shows the progress as days, hours, minutes and seconds elapsed, counting up from 0.
Example (<1 day): “T+01:30:25” (i.e. 1 hour, 30 minutes, 25 seconds elapsed)
Example (>=1 day): “T+1d30:25” (i.e. 1 day, 30 hours, 25 minutes elapsed)
REM
Shows the progress as days, hours, minutes, and seconds remaining, counting down to 0.
Example (<1 day): “T–01:30:25” (i.e. 1 hour, 30 minutes, 25 seconds remaining)
Example (>=1 day): “T–1d30:25” (i.e. 1 day, 30 hours, 25 minutes remaining)
To select a PROGRESS report mode, press:
STEPS Submenu
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
Make sure that the M703E
is in standby mode.
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT
SETUP X.X SEQUENCE CONFIGURATION
EDIT PRINT EXIT
SETUP X.X END OF SEQUENCES
INS PRNT EXIT
SETUP X.X 1) SEQ [NAME], [X] STEPS
PREV NEXT INS DEL EDIT PRNT EXIT
SETUP X.X NAME:0
SET> EDIT EXIT
Deletes the sequence shown
in the message field
Edits the sequence shown
in the message field
Scrolls back and forth between
existing sequences
Continue pressing SET> until ...
This display only appears if there are no sequences currently
programmed into the M703E.
OTHERWISE ...
SETUP X.X PROGRESS MODE:REM
<SET EDIT EXIT
SETUP X.X PROGRESS MODE:REM
STEP PCT ELAP REM ENTR EXIT
Use these keys to choose
a PROGRESS MODE
8
SETUP X.X PROGRESS MODE:ELAP
<SET EDIT EXIT
05744 Rev B 49