Teledyne 703E Photo Scanner User Manual


 
TELEDYNE API
Operating the M703E Calibrator M703E Calibrator Operator’s Manual
70 05744 Rev B
6.9.2.2. Automatic TEST CHANNEL Calibration
To calibrate the outputs as a group with the AOUTS CALIBRATION command, press:
NOTE
Before performing this procedure, make sure that the AUTO CAL feature is turned OFF for CONC_OUT_1
and CONC_OUT_2,
Make sure that the AUTO CAL feature is turned ON for the TEST CHANNEL (See Section 6.9.2.1)
From the
AIO CONFIGURATION SUBMENU
(See Section 6.9.1.1.)
Analyzer
automatically
calibrates all
channels for which
AUTO-CAL is turned
ON
If any of the channels have not
been calibrated ot if at least one
channel has AUTO-CAL turned
OFF, this message will read NO.
DIAG AIO NOT AUTO CAL. CONC_OUT_1
DIAG AIO NOT AUTO CAL. CONC_OUT_2
DIAG AIO AUTO CALIBRATING TEST_OUTPUT
This message
appears when
AUTO-CAL is
Turned OFF for
a channel
DIAG AIO AOUTS CALIBRATED: YES
SET> CAL EXIT
DIAG AIO AOUTS CALIBRATED: NO
SET> CAL EXIT
DIAG ANALOG I/O CONFIGURATION
PREV NEXT ENTR EXIT
NOTE:
Manual calibration sho tputs must be closely uld be used for the 0.1V range or in cases where the ou
matched to the characteristics of the recording device.