M703E Calibrator Operator’s Manual General Troubleshooting & Repair of the M703E Calibrator
11.4.8.4. Control Inputs
Table 11-11: M703E Control Input Pin Assignments and Corresponding Signal I/O Functions
CONNECTOR INPUT CORRESPONDING I/O SIGNAL
Top A
CONTROL_IN_1
Top B
CONTROL_IN_2
Top C
CONTROL_IN_3
Top D
CONTROL_IN_4
Top E
CONTROL_IN_5
Top F
CONTROL_IN_6
Bottom G
CONTROL_IN_7
Bottom H
CONTROL_IN_8
Bottom I
CONTROL_IN_9
Bottom J
CONTROL_IN_10
Bottom K
CONTROL_IN_11
Bottom L
CONTROL_IN_12
The control input bits can be tested by applying a trigger voltage to an input and watching changes in the status
of the associated function under the SIGNAL I/O submenu:
EXAMPLE: to test the “
A” control input:
1. Under the
DIAG SIGNAL I/O menu (See Section11.1.3), scroll through the inputs and outputs until
you get to the output named
0) CONTROL_IN_1.
2. Connect a jumper from the
+” pin on the appropriate connector to the “U” on the same connector.
3. Connect a second jumper from the “” pin on the connector to the “
A” pin.
4. The status of
0) CONTROL_IN_1 should change to read “ON”.
05744 Rev B 193