TELEDYNE API
Operating the M703E Calibrator M703E Calibrator Operator’s Manual
6.5.1.2. Naming a Sequence
The first step of creating a calibration sequence is to assign it a name. The name can be up to 10 characters
and can be comprised of any alpha character (A to Z), and numeral (0 to 9) or the underscore character (“_“).
To assign a name to a sequence, press,
SETUP X.X NAME:[0]
<CH CH> INS DEL [0] ENTER EXIT
EXIT discards the
new NAME
ENTR accepts the
new NAME
Toggle this key to cycle
through the range of
numerals and available
characters:
(“A – Z”; “0 – 9” & “ _ ”)
Moves the
cursor one
character left or
right.
Deletes a
character at the
cursor location.
Inserts a new a
character at the
cursor location.
STANDBY ACT CAL=0.000 LPM
<TST TST> GEN STBY SEQ SETUP
Make sure that the M703E
is in standby mode.
SETUP X.X PRIMARY SETUP MENU
GAS SEQ CFG CLK PASS MORE EXIT
SETUP X.X SEQUENCE CONFIGURATION
EDIT PRINT EXIT
SETUP X.X END OF SEQUENCES
INS PRNT EXIT
SETUP X.X 1) SEQ [NAME], [X] STEPS
PREV NEXT INS DEL EDIT PRNT EXIT
SETUP X.X NAME:0
SET> EDIT EXIT
Deletes the sequence shown
in the message field
Edits the sequence shown
in the message field
Scrolls back and forth between
existing sequences
This display only appears if there are no sequences currently
programmed into the M703E.
OTHERWISE ...
42 05744 Rev B