TELEDYNE API
Operating The M703E over the Serial I/O Ports M703E Calibrator Operator’s Manual
7.5.2.1. Manually Configuring the Network IP Addresses
There are several circumstances when you may need to configure the interface settings of the calibrator’s
Ethernet card manually. The
INET sub-menu may also be used to edit the Ethernet card’s configuration
properties
Your LAN is not running a DHCP software package,
The DHCP software is unable to initialize the calibrator’s interface;
You wish to program the interface with a specific set of IP addresses that may not be the ones
automatically chosen by DHCP.
Editing the Ethernet Interface properties is a two-step process.
STEP 1: Turn DHCP OFF: While DHCP is turned ON, the ability to set INSTRUMENT IP, GATEWAY IP and
SUBNET MASK manually is disabled
SETUP X.X COMMUNICATIONS MENU
ID ADDR INET EXIT
SETUP X.X ENTER PASSWORD:818
81 8 ENTREXIT
SETUP X.X DHCP:ON
<SET SET> EDIT EXIT
SETUP X.X DHCP:ON
ON ENTR EXIT
SETUP X.X DHCP:OFF
OFF ENTR EXIT
Toggle this key
to turn
DHCP
ON/OFF
ENTR accepts
the new setting
EXIT ignores the
new setting
Continue to Step 2 Below
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT