M703E Calibrator Operator’s Manual Getting Started
3.4. INITIAL OPERATION
If you are unfamiliar with the M703E theory of operation, we recommend that you read Chapter 9.
For information on navigating the calibrator’s software menus, see the menu trees described in Appendix A.1.
3.4.1. START-UP
After all of the electrical and pneumatic connections are made, turn on the instrument. The exhaust fan and
should start immediately. If the instrument is equipped with an internal photometer installed, the associated
pump should also start up.
The display should immediately display a single, horizontal dash in the upper left corner of the display. This will
last approximately 30 seconds while the CPU loads the operating system.
Once the CPU has completed this activity, it will begin loading the calibrator firmware and configuration data.
During this process, string of messages will appear on the calibrator’s front panel display:
Firmware fully
booted
The revision level of the
firmware installed in your
analyzer is briefly displayed
SOFTWARE REVISION B.0
BOOT PROGRESS [XXXXXXXX 80% _ _]
The instrument is loading
the calibrator firmware.
M703E CALIBRATOR
BOOT PROGRESS [XXXXX 50%_ _ _ _ _]
The instrument is loading
configuration and calibration
data from the flash chip
STARTING INSTRUMENT W/FLASH : 1
System waits 3 seconds
then automatically begins its
initialization routine.
No action required.
System is checking the format of
the instrument’s flash memory
chip.
SELECT START OR REMOTE : 3
START .
CHECKING FLASH STATUS : 1
STARTING INSTRUMENT CODE : 1
STANDBY SYSTEM RESET
TEST GEN STBY SEQ MSG CLR SETUP
If at this point,
**FLASH FORMAT INVALID**
appears, contact T–API customer service
Press CLR to clear initial
warning messages.
The calibrator should automatically switch to STANDBY mode after completing the boot-up sequence.
05744 Rev B 21