M703E Calibrator Operator’s Manual Operating the M703E Calibrator
05744 Rev B 83
6.12.5.2. Creating a MANUAL LEVEL
To create a level using the M703E’s MANUAL generation function, press:
SETUP X.X INSERT STEP: MANUAL
PREV ENTR EXIT
CHOOSE ACTION Submenu
SETUP X.X ACTION TO PERFORM:GENERATE
NEXT ENTR EXIT
Use the NEXT until ...
Starting at the CHOOSE ACTION Submenu
SETUP X.X O3 GEN SET POINT: 000.0 PPB
0 0 0 0 .0 ENTR EXIT
SETUP X.X O3 GEN MODE: OFF
OFF CNST REF BNCH ENTR EXIT
This Key
Turns the the
O
3
Generator
OFF/ON
REF: The concentration control
loop will use the generator’s
reference detector as input.
BNCH: The concentration
control loop will use the
photometer bench.
SETUP X.X O3 GEN SET POINT: 0.0 MV
0 0 0 0 .0 ENTR EXIT
This key sets a CONSTANT drive
voltage for the O
3
Generator
These keys set a target
concentration for the O
3
Generator
Toggle these keys
to set the
CONSTANT drive
voltage of the O
3
generator
Toggle these keys
to set output
CONCENTRATION
of the O
3
generator
EXIT discards the new
setting
ENTR accepts the
new setting
STANDBY LEVEL:0
00 ENTREXIT
Toggle these keys until
the designation of the
existing defined level
program is reached.
EXIT discards the new
LEVEL number
ENTR accepts the new
LEVEL number