Teledyne 703E Photo Scanner User Manual


 
TELEDYNE API
M703E Calibrator Operator’s Manual Operating the M703E Calibrator
Example: If all passwords are enabled, the following keypad sequence would be required to enter the VARS or
DIAG submenus:
SYSTEM ENTER SETUP PASS:0
8 1 8 ENTREXIT
SYSTEM ENTER SETUP PASS:0
0 0 0 ENTR EXIT
SETUP X.X PRIMARY SETUP MENU
O3 SEQ CFG CLK PASS MORE EXIT
SETUP X.X SECONDARY SETUP MENU
COMM VARS DIAG EXIT
EXAMPLE: This
password enables the
SETUP mode
Press individual
keys to set
number
M703E enters selected menu
STANDBY ACT =STANDBY
<TST TST> GEN STBY SEQ SETUP
NOTE
The instrument still prompts for a password when entering the VARS and DIAG menus, even if
passwords are disabled, but it displays the default password (818) upon entering these menus. The
user only has to press ENTR to access the password-protected menus but does not have to enter the
required number code.
05744 Rev B 61