Teledyne 703E Photo Scanner User Manual


 
TELEDYNE API
Operating the M703E Calibrator M703E Calibrator Operator’s Manual
6.9.2.3. Manual Calibration of the TEST CHANNEL configured for Voltage Ranges
For highest accuracy, the voltages of the analog outputs can be manually calibrated.
NOTE:
The menu for manually adjusting the analog output signal level will only appear if the AUTO-CAL feature
is turned off for the channel being adjusted (see Section6.9.2.1)
Calibration is performed with a voltmeter connected across the output terminals (See Figure 6-2) and by
changing the actual output signal level using the front panel keys in 100, 10 or 1 count increments.
V
+DC Gnd
Figure 6-3: Setup for Calibrating the TEST CHANNEL
Table 6-10: Voltage Tolerances for the TEST CHANNEL Calibration
FULL
SCALE
ZERO
TOLERANCE
SPAN VOLTAGE
SPAN
TOLERANCE
MINIMUM
ADJUSTMENT
(1 count)
0.1 VDC ±0.0005V 90 mV ±0.001V 0.02 mV
1 VDC ±0.001V 900 mV ±0.001V 0.24 mV
5 VDC ±0.002V 4500 mV ±0.003V 1.22 mV
10 VDC ±0.004V 4500 mV ±0.006V 2.44 mV
72 05744 Rev B